Web Sensors (TAx6x0) provide temperature, relative humidity, barometric pressure, and CO2 concentration measurements using internal sensors. Values are shown on the device display and are accessible via an Ethernet web interface, with optional notification and cloud/database data forwarding.
Spesifikasi
| Jenis peranti | Web Sensor (built-in internal sensors) |
|---|---|
| model | TA0610, TA3610, TA5640, TA7610, TA7640 |
| Measured/computed values (as listed) | TA0610: T TA3610: T + RH + CV TA5640: P + CO2 TA7610: T + RH + P + CV TA7640: T + RH + P + CO2 + CV |
| Bekalan voltage (coaxial connector, 5.1 x 2.1 mm) | 5 to 24 Vdc (all listed models) |
| Kuasa ke atas Ethernet | IEEE 802.3af, PD Class 0 (max. 12.95 W), voltage from 36 V to 57 Vdc |
| Penggunaan kuasa | Approx. 1 W (all listed models) |
| Julat pengukuran suhu | -30 to +60°C (TA0610/TA3610/TA7610/TA7640); TA5640: — |
| Ketepatan suhu | ± 0.4°C (where temperature is measured: TA0610/TA3610/TA7610/TA7640) |
| Julat kelembapan relatif | 0 to 100 %RH (TA3610/TA7610/TA7640); TA0610/TA5640: — |
| Relative humidity accuracy (5 to 95 %RH at 23°C) | ± 2.5 %RH (where RH is measured: TA3610/TA7610/TA7640) |
| Julat tekanan barometrik | 600 to 1100 hPa (TA5640/TA7610/TA7640); TA0610/TA3610: — |
| Barometric pressure accuracy (at 23°C) | ± 1.3 hPa (where pressure is measured: TA7610/TA5640/TA7640) |
| CO2 concentration range | 400 to 10,000 ppm (TA5640/TA7640); TA0610/TA3610/TA7610: — |
| CO2 accuracy (400 to 5,000 ppm) | ±(50 ppm + 3% of measured value) (models with CO2) |
| Selang penentukuran yang disyorkan | TA0610: 2 years; TA3610: 1 year; TA5640: 5 years; TA7610: 1 year; TA7640: 1 year |
| Kelas perlindungan | IP30 (all listed models) |
| Operating temperature/storage temperature | -30 to +60°C (all listed models) |
| Operating humidity (no condensation) | TA0610/TA3610: 0 to 100 %RH TA5640/TA7610/TA7640: 0 to 85 %RH |
| Kelembapan operasi/penyimpanan | Storage humidity: 0 to 85 %RH (all listed models) |
| Kedudukan pemasangan | Sensor cover downwards (TA0610/TA3610/TA7610/TA7640); TA5640: any position |
| Berat badan | TA0610: 220 g; TA3610: 230 g; TA5640: 210 g; TA7610: —; TA7640: 240 g |
Quick setup and use
- Select a holder mounting method, fasten the holder, and slide the device onto the holder.
- Connect cables: connect the Ethernet cable (or connect the power adapter if PoE is not used).
- Ensure you have a free IP address for the device (DHCP is default). After connecting, determine the device IP address using one of: COMET Vision search, checking the gateway/DHCP server configuration, or pressing the device button to display the IP address (if acoustic signaling is not active).
- Masukkan alamat IP peranti ke dalam a web pelayar untuk mengakses web antara muka.
- To configure the device, activate configuration mode by pressing the button on the home page, then save changes (changes are applied after saving).
PENERANGAN PRODUK
- Web Sensors TAx6x0 with an Ethernet connection are compact devices featuring internal sensors for temperature, relative humidity, barometric pressure, and CO2 concentration measurement. The devices can be powered either by a power supply adapter or via Power over Ethernet (PoE).
- The kepekatan CO2 is measured by an NDIR sensor using LED technology, ensuring low energy consumption and long-term measurement stability.
- Sensors with relative humidity measurement can also calculate derived humidity quantities such as dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy, humidex, and heat index.
- Nilai yang diukur dan dikira are displayed on a two-line LCD or can be read and processed via the Ethernet connection. The values are available on the device’s web interface and can be sent to the COMET Cloud online portal or the local COMET Database storage. The sensors also support integration with third-party data collection systems via Modbus TCP, HTTP GET (JSON and XML), HTTP POST (JSON), and SNMP protocols. An important feature of the devices is the ability to check measured values and notify via warning messages or acoustic/optical signals. Notification methods include status indication on the device’s website, e-mail alerts, and the Syslog protocol. The device can be configured through its web interface or via COMET Vision software, which is available free of charge from the manufacturer’s webtapak (www.cometsystem.com).
| taip | nilai yang diukur | versi | melekap |
|---|---|---|---|
| TA0610 | T | udara persekitaran | dinding |
| TA3610 | T + RH + CV | udara persekitaran | dinding |
| TA5640 | P + CO2 | udara persekitaran | dinding |
| TA7610 | T + RH + P + CV | udara persekitaran | dinding |
| TA7640 | T + RH + P + CO2 + CV | udara persekitaran | dinding |
T…temperature, RH…relative humidity, CO2…concentration of CO2 in air, P…barometric pressure, CV…computed values
PEMASANGAN DAN PENGENDALIAN
The devices should be mounted on a flat surface according to the procedure detailed below. Careful attention should be paid to the installation of the devices and probes. Incorrect positioning or placement can adversely affect the accuracy and long-term stability of the measured values. Devices with a measuring stem must always be installed with the sensor cover facing downwards.
The devices do not require any special operation or maintenance. Periodic calibration is recommended to ensure measurement accuracy.
PENYEDIAAN PERANTI
To install the device and connect it to the network, you must have a free IP address. This address can be either static or obtained automatically from a DHCP server. Contact your network administrator and request permission before connecting the device. The device supports both IPv4 and IPv6 protocols. The device is configured to use DHCP by default. The IPv6 protocol is enabled in automatic mode (DHCPv6 or SLAAC) by default.
Connect the network cable and, if necessary, plug in the power adapter. Since the device is set to DHCP by default, you need to determine its IP address after connecting it. This can be done by searching for the device using the COMET Vision program, viewing the IP address in the gateway/DHCP server configuration, or displaying the IP address on the device’s display by pressing the button. Then enter the IP address into your web browser. From there, you can change the device settings. Configuration mode is activated by pressing the button on the home
page. The changes are applied after saving the configuration.
PROTOKOL KOMUNIKASI DAN NEGERI RALAT
A detailed description of communication protocols and error states is available in the user manual on the manufacturer’s website. A brief overview of the communication protocols can also be viewed directly on the device’s web antara muka.
ARAHAN KESELAMATAN
- The device does not have enhanced protection against water or dust ingress. Never use it in environments with splashing or dripping water, condensation, aerosols, or excessive dust.
- The device must not be operated or stored without a sensor cover.
- The humidity sensors should not be exposed to water or other liquids.
- Use a power adapter approved according to the relevant standards. Use only IEEE 802.3af PoE.
- The device is not intended for use in critical applications where a communication failure could endanger health or cause material damage. Ensure proper cybersecurity measures when operating the device.
- Installation should only be performed by qualified personnel in accordance with applicable regulations and standards.
- The devices contain electronic components and must be disposed of in accordance with current regulations.
- Untuk maklumat tambahan, refer to the manuals and other documentation available in the “Download” section for each device at www.cometsystem.com
Spesifikasi teknikal

* valid for the TA7610 device only
** valid at a temperature of 15-35°C. In the measurement range of 5 000 to 10 000 ppm, the accuracy is ±10% of the measured value.
Button – functions

- Factory defaults procedure: turn off the device • press and hold the button • turn on the device • release the button
- Mute of active acoustic signaling: tekan butang sebentar
- Displaying device IP address (in case the acoustic signaling is not active): press the button briefly
Pemasangan peranti
- Select holder mounting method: Select one of the device holder mounting methods and fasten the holder.

- Slide device onto holder: Slide the device onto the holder.

- Sambungkan kabel: Connect the Ethernet cable (if PoE power supply is not used, connect the adapter).

Menyelesaikan masalah
- Communication/protocol issues: A detailed description of communication protocols and error states is available in the user manual on the manufacturer’s website. A brief overview is also available directly on the device’s web antara muka.
- Cannot find the device IP address: Because DHCP is used by default, determine the IP address after connecting via COMET Vision, the gateway/DHCP server configuration, or by pressing the device button to display the IP address (when acoustic signaling is not active).
For additional issues not covered above, refer to the full manual available in the manufacturer’s download section at www.cometsystem.com.