HACH 146900

Kit Ujian Oksigen Terlarut Hach 146900, Model OX-2P

Manual Pengguna

1. Pengenalan

The Hach 146900 Dissolved Oxygen Test Kit, Model OX-2P, is designed for the accurate determination of dissolved oxygen concentration in water samples. This kit utilizes a simple and economical drop count titration method, making it suitable for field and laboratory use. It provides all necessary reagents and apparatus for approximately 100 tests, conveniently packaged in a durable plastic carrying case.

2. Komponen dan Persediaan Kit

Before beginning any tests, ensure all components are present and in good condition. Familiarize yourself with each item in the kit.

Hach 146900 Dissolved Oxygen Test Kit components laid out

Rajah 2.1: Berakhirview of the Hach 146900 Dissolved Oxygen Test Kit components. This image displays the complete kit contents, including the carrying case, various reagent bottles, powder packets, a test tube, and other necessary apparatus for conducting dissolved oxygen measurements.

Hach 146900 Dissolved Oxygen Test Kit opened, showing reagents

Rajah 2.2: The Hach 146900 Dissolved Oxygen Test Kit with the lid open, revealing the organized storage of reagents and testing apparatus. This view highlights the specific reagent bottles and packets, such as PERMACHEM reagents, stored within the kit's carrying case.

Kit biasanya termasuk:

  • Sarung Pembawa Plastik
  • Reagent Bottles (e.g., Manganous Sulfate Solution, Alkaline Iodide-Azide Reagent, Sulfuric Acid)
  • PERMACHEM Reagent Powder Pillows
  • Titration Apparatus (e.g., Titration Tube, Measuring Tube)
  • Dropper Bottle for Titrant
  • Sample Bottles

Ensure all bottles are securely capped and powder pillows are sealed. Store the kit in a cool, dry place away from direct sunlight.

3. Operating Instructions (Drop Count Titration Method)

This section outlines the general procedure for determining dissolved oxygen using the drop count titration method. Always refer to the specific instructions provided with your reagent lot for precise details, as procedures may vary slightly.

3.1. Sample Koleksi

  1. Kumpul air sample carefully to avoid aeration. Fill the sample bottle completely, allowing some overflow, then cap it without trapping air bubbles.
  2. Perform the test immediately after sample collection for most accurate results, especially for unstable samples.

3.2. Reagent Addition and Fixation

  1. Buka sample bottle and add the specified number of drops of Manganous Sulfate Solution.
  2. Immediately add the specified number of drops of Alkaline Iodide-Azide Reagent.
  3. Cap the bottle and invert several times to mix. A flocculent precipitate will form.
  4. Allow the precipitate to settle for a few minutes until a clear supernatant forms above the precipitate.
  5. Add the specified amount of Sulfuric Acid (or other acid reagent as per instructions). Cap and invert to mix until the precipitate dissolves, forming a yellowish-brown solution. This step fixes the oxygen.

3.3. Titration

  1. Transfer a measured volume of the fixed sample (e.g., 20 mL) into the titration tube.
  2. Add the specified PERMACHEM Reagent Powder Pillow contents to the sample in the titration tube. Mix to dissolve.
  3. Slowly add the titrant (e.g., Sodium Thiosulfate Solution) drop by drop from the dropper bottle, swirling the tube after each drop.
  4. Continue adding drops until the yellowish-brown color just disappears and the solution becomes colorless. This is the endpoint.
  5. Count the number of drops used. Each drop typically corresponds to a specific concentration of dissolved oxygen (e.g., 1 drop = 1 mg/L DO). Refer to your kit's specific conversion factor.

3.4. Tafsiran Keputusan

The number of drops of titrant used directly corresponds to the dissolved oxygen concentration in mg/L (parts per million, ppm). Consult the specific instructions or conversion chart provided with your kit for accurate interpretation.

4. Penyelenggaraan dan Penyimpanan

  • Pembersihan: Rinse all glassware and plasticware thoroughly with distilled or deionized water after each use. Allow to air dry.
  • Penyimpanan Reagen: Keep all reagent bottles tightly capped when not in use to prevent contamination and degradation. Store the entire kit in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Tarikh Luput: Pay attention to the expiration dates on reagents. Expired reagents can lead to inaccurate results. Dispose of expired reagents according to local regulations.
  • Kit Integrity: Periodically inspect the carrying case for damage. Ensure all components are securely stored to prevent breakage or spills during transport.

5. Penyelesaian masalah

MasalahKemungkinan PuncaPenyelesaian
Keputusan yang tidak tepat atau tidak konsistenReagen yang telah tamat tempoh, bahan yang tidak betulample collection (aeration), contaminated glassware, incorrect procedure steps.Check reagent expiration dates. Ensure proper sample collection technique. Thoroughly clean all glassware. Review and follow the operating instructions carefully.
No precipitate forms after reagent additionIncorrect reagent addition order or quantity, reagents degraded.Verify reagent addition steps and quantities. Ensure reagents are not expired or contaminated.
Color does not disappear during titrationInsufficient titrant, titrant degraded, very high DO concentration.Ensure titrant is not expired. If DO is extremely high, dilute the sample and re-test (adjusting calculation).
Kit components missing or damagedShipping damage, improper storage, loss during use.Contact Hach customer support for replacement parts. Inspect kit upon receipt.

6. Spesifikasi

AtributPerincian
Nombor Model146900
JenamaHATI
Kaedah UjianDrop Count Titration
Bilangan UjianLebih kurang 100
Dimensi Pakej11.7 x 7.7 x 5.8 inci
Berat Barang3.1 paun
PengeluarHach
Tarikh Tersedia Pertama28 Ogos 2014

7. Waranti dan Sokongan

Specific warranty information for the Hach 146900 Dissolved Oxygen Test Kit is typically provided at the point of purchase or within the product packaging. For detailed warranty terms, technical support, or to order replacement parts and reagents, please contact Hach customer service directly or visit their official webtapak.

Hach Company Contact Information: Please refer to the contact details provided on the Hach official website or product packaging for the most current support options.

Dokumen Berkaitan - 146900

Praview Manual Pengguna Sensor Oksigen Terlarut Bercahaya Intelek HACH LBOD101
Manual pengguna untuk prob HACH Intellical LBOD101, sensor oksigen terlarut pendarkilau digital yang direka untuk mengukur kepekatan oksigen terlarut dalam BODamples. Panduan ini merangkumi produk di seluruhview, specifications, safety information, preparation, calibration procedures (100% air-saturated and zero-point), sample measurement, maintenance, troubleshooting, and consumables.
Praview Hach Polymetron 9582 Dissolved Oxygen System with HART Communications
Detailed specifications and features of the Hach Polymetron 9582 Dissolved Oxygen System, a 24 V DC instrument with HART communications for water analytics in the power industry. Includes information on design, installation, operation, maintenance, and technical specifications.
Praview Manual Pengguna Prob Oksigen Terlarut Bercahaya Siri Hach Intellical LDO101
Manual pengguna untuk prob oksigen terlarut pendarkilau digital siri Hach Intellical LDO101. Meliputi produk di seluruhview, spesifikasi, keselamatan, penyediaan, penentukuran, pengukuran, penyelenggaraan, penyelesaian masalah dan bahan habis pakai.
Praview Manual Arahan Meter Oksigen Terlarut Mudah Alih Hach session6
Manual arahan komprehensif untuk Meter Oksigen Terlarut Mudah Alih Hach session6, meliputi persediaan, operasi, penentukuran, penyelenggaraan dan penyelesaian masalah untuk pengukuran oksigen terlarut yang tepat dalam larutan akueus.
Praview Manual Pengguna Hach LDO Sensor
Manual pengguna ini menyediakan arahan komprehensif untuk Sensor LDO Hach, meliputi spesifikasi, pemasangan, operasi, penentukuran dan penyelesaian masalah untuk aplikasi air sisa perbandaran dan industri.
Praview Polymetron 9582 Dissolved Oxygen System with MODBUS Communications
Technical specifications and features of the Hach Polymetron 9582 Dissolved Oxygen System, a component of the Hach water analytics system for the Power industry. Includes details on design, installation, operation, maintenance, and technical specifications.