1. Pengenalan
The Extech PH110 Waterproof ExStik pH Meter is designed for accurate pH measurements in various substances. This device features a refillable pH electrode, allowing for extended use and consistent accuracy by replenishing the reference solution. Its robust, waterproof design ensures reliability in diverse environments.
2. Ciri-ciri Utama
- Refillable pH electrode for extended service life.
- Analog bargraph display on LCD for trend indication.
- Waterproof construction to IP57 standard.
- Automatic 1, 2, or 3 point calibration with buffer solution recognition.
- Flat Surface Electrode capable of measuring pH in liquids, semi-solids, and solids.
3. Kandungan Pakej
The Extech PH110 Waterproof ExStik pH Meter package includes:
- Extech PH110 pH Meter with refillable pH electrode
- 1.5mL bottle of KCI (Potassium Chloride) solution
- Two reference junctions
- Two electrode removal tools
- Tudung pelindung
- Sample cawan
- Four 3V CR-2032 batteries (pre-installed or included separately)
- 48-inch (1.2m) neck strap
4. Persediaan
4.1 Pemasangan Bateri
- Ensure the meter is off.
- Cari petak bateri.
- Insert four 3V CR-2032 Lithium Metal batteries, observing correct polarity.
- Tutup petak bateri dengan selamat.
4.2 Penyediaan Elektrod
- Tanggalkan penutup pelindung dari elektrod.
- Rinse the electrode tip with distilled water.
- If the electrode appears dry, soak it in electrode storage solution or pH 7 buffer for at least 30 minutes before first use.
5. Operasi
5.1 Penentukuran
Calibration is essential for accurate measurements. The PH110 supports 1, 2, or 3 point calibration and automatically recognizes buffer solutions.
- Turn on the meter by pressing the HIDUP/MATI butang.
- Prepare fresh pH buffer solutions (e.g., pH 4, 7, and 10). These are sold separately.
- Tekan dan tahan butang CAL button until 'CAL' appears on the display.
- Immerse the electrode into the first buffer solution (e.g., pH 7). The meter will automatically recognize the buffer and calibrate. Wait for the reading to stabilize.
- For 2 or 3 point calibration, rinse the electrode with distilled water and immerse it into the next buffer solution (e.g., pH 4, then pH 10). The meter will continue to calibrate automatically.
- Ensure the display shows 'CAL' during the calibration process.
5.2 Pengukuran
The flat surface electrode allows for versatile measurements.
- After calibration, rinse the electrode with distilled water.
- Immerse the electrode into the sample to be measured. For liquids, ensure the electrode tip is fully submerged. For semi-solids or solids, ensure good contact with the flat surface electrode.
- Allow the reading to stabilize on the display. The analog bargraph will indicate trends in acidity or alkalinity.
- To switch between pH and temperature display modes, press the MODE butang.
- After measurement, rinse the electrode thoroughly with distilled water and replace the protective cap.

Rajah 1: The Extech PH110 pH meter, a handheld device with a digital display showing pH and temperature readings. It features three control buttons: MODE, CAL, and ON/OFF.
6. Penyelenggaraan
6.1 Electrode Refilling
The refillable electrode design allows for prolonged accuracy and lifespan.
- When the reference solution depletes or accuracy declines, use the provided removal tools to access the reference junctions.
- Carefully refill the electrode with the 1.5mL KCI solution.
- Ensure the reference junctions are properly re-seated after refilling.

Rajah 2: Satu jarak dekat view of hands refilling the Extech PH110 pH meter's electrode module with a specialized electrode filling solution from a small bottle. This process ensures the meter's continued accuracy and extends electrode life.
6.2 Pembersihan dan Penyimpanan
- Always rinse the electrode with distilled water after each use to prevent contamination.
- The flat surface electrode is designed for easy cleaning. Gently wipe with a soft cloth if necessary.
- Store the electrode with its protective cap filled with electrode storage solution (or pH 7 buffer if storage solution is unavailable) to keep the sensor hydrated. Do not store dry.
7. Penyelesaian masalah
If the meter exhibits inaccurate readings or operational issues, consider the following:
- Bacaan Tidak Tepat: Recalibrate the meter using fresh buffer solutions. Ensure the electrode is clean and properly hydrated. Check if the reference solution in the electrode needs refilling.
- Tiada Isu Paparan/Kuasa: Verify that the batteries are correctly installed and have sufficient charge. Replace batteries if necessary.
- Respons Perlahan: The electrode might be dry or contaminated. Clean and rehydrate the electrode, then recalibrate.
8. Spesifikasi
| Spesifikasi | Perincian |
|---|---|
| Pengeluar | Extech |
| Nombor Bahagian | PH110 |
| Berat Barang | 3.87 auns |
| Dimensi Produk | 6.8 x 1.6 x 1.4 inci |
| Nombor Model Item | PH110 |
| Bateri | 4 Lithium Metal batteries (CR-2032) required (included) |
| Sumber Kuasa | Dikuasakan Bateri |
| Penilaian Kalis Air | Piawaian IP57 |
| Penentukuran | 1, 2, or 3 point automatic buffer recognition |
| Jenis Elektrod | Refillable Flat Surface Electrode |
9. Maklumat Waranti
Extech provides a full one-year warranty on parts and workmanship for the PH110 pH Meter, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modifications.
10. Sokongan
For further assistance, technical support, or to purchase accessories and replacement parts, please refer to the official Extech Instruments weblaman web atau hubungi jabatan khidmat pelanggan mereka. Maklumat hubungan biasanya boleh didapati pada pembungkusan pengeluar atau webtapak.