Quantek Instruments 901 Headspace Oxygen Analyzer User Manual

pengenalan

The Quantek Instruments 901 Headspace Oxygen Analyzer is a compact, handheld device designed for measuring residual oxygen levels in sealed packages. Its portable design allows for on-site analysis, providing accurate readings for quality control and process monitoring. This manual provides essential information for the proper setup, operation, maintenance, and troubleshooting of your analyzer.

Notis Keselamatan Penting: This product is not approved or intended for, and should not be used for medical, clinical, surgical, or other patient-oriented applications.

Produk Selesaiview

Quantek Instruments 901 Headspace Oxygen Analyzer with sampling kuar

Image: The Quantek Instruments 901 Headspace Oxygen Analyzer, a black and blue handheld device with a digital display showing "21.0 PERCENT O₂", two control buttons labeled "POWER" and "PUMP", and indicator lights for "Low Batt" and "Sampling". A white sampling probe with a thin tube extends from the top right.

1. Persediaan

1.1 Membongkar dan Pemeriksaan

Keluarkan semua komponen dari pembungkusan dengan berhati-hati. Sahkan bahawa semua item yang disertakan ada:

1.2 Pemasangan dan Pengecasan Bateri

  1. Open the battery compartment located on the back of the analyzer.
  2. Insert the four rechargeable NiCad batteries, ensuring correct polarity.
  3. Tutup petak bateri dengan selamat.
  4. Connect the charger to the analyzer and plug the 115 VAC adapter into a power outlet. Allow the batteries to charge fully before first use.

Nota: The analyzer must always be used with rechargeable NiCad batteries.

1.3 Attaching the Sampling Probe

  1. Carefully attach one of the needles to the sampling probe assembly.
  2. Sambungkan sampling probe assembly to the designated port on the top of the analyzer. Ensure a secure connection to prevent leaks.
  3. Install a probe filter into the sampling probe assembly.

2. Arahan Operasi

2.1 Menghidupkan/Mematikan

2.2 Measuring Oxygen Levels

  1. Ensure the analyzer is powered on and the sampling probe is correctly attached with a fresh needle and filter.
  2. Place a foam seal pad on the package surface where you intend to sample.
  3. Carefully pierce the package through the foam seal pad with the sampling needle.
  4. Tekan ke PAM button to activate the internal pump. The "Sampling" indicator light will illuminate. The pump will draw a sample from the package headspace to the sensor.
  5. Observe the digital display for the oxygen percentage reading. Allow the reading to stabilize.
  6. Once the reading is stable, release the PAM butang.
  7. Carefully withdraw the needle from the package.

The analyzer is capable of measuring oxygen levels from very low to very high concentrations (50% to 100%) without sensor overload.

3. Penyelenggaraan

3.1 Probe Filter Replacement

Regularly inspect and replace the probe filter to ensure accurate readings and prevent contamination of the sensor. Replace the filter if it appears discolored or clogged.

3.2 Pembersihan

3.3 Penjagaan Bateri

4. Penyelesaian masalah

MasalahKemungkinan PuncaPenyelesaian
Penganalisis tidak boleh dihidupkan.Bateri mati atau tidak dipasang dengan betul.Check battery installation and charge the batteries. Ensure only NiCad batteries are used.
"Low Batt" indicator is on.Batteries require charging.Connect the analyzer to the charger and allow batteries to fully charge.
Bacaan yang tidak tepat atau turun naik.Clogged probe filter, loose probe connection, or sensor issue.Replace the probe filter. Check that the sampling probe is securely connected. If issues persist, contact support.
Pump does not activate.Bateri lemah atau kerosakan dalaman.Ensure batteries are charged. If the problem continues, contact support.

5. Spesifikasi

6. Waranti dan Sokongan

For warranty information, technical support, or service inquiries, please contact your authorized Quantek Instruments dealer or visit the official Quantek Instruments website. Please have your model number (AO-10360-20) and purchase date available when contacting support.

Dokumen Berkaitan - AO-10360-20

Praview Quantek KPS Access Control Keypad & Reader User Manual
Comprehensive user manual for the Quantek KPS Access Control Keypad & Reader, detailing installation, programming, features, specifications, and troubleshooting for secure door access.
Praview Manual Pengguna Quantek CP5-RX: Pad Kekunci Jarak dan Pembaca Wiegand
Manual pengguna untuk Quantek CP5-RX, papan kekunci keluaran Wiegand dengan pembaca jarak bersepadu. Meliputi spesifikasi, pemasangan, pendawaian, pengaturcaraan, isyarat data dan format penghantaran pad kekunci untuk kad EM dan Mifare.
Praview Manual Pengguna Pengawal Akses Pelbagai fungsi KPFA-BT
Manual pengguna untuk pengawal capaian berbilang fungsi KPFA-BT dengan pengaturcaraan Bluetooth. Ketahui tentang spesifikasinya, pendawaian, pengendalian aplikasi, menghantar eKey, menjana kod laluan, menambah kad, cap jari, alat kawalan jauh, mengurus pentadbir yang dibenarkan dan viewrekod.
Praview CP-611LSR-V2.0 Laser Photocell Sensor - Specifications and Setup Guide
Detailed technical specifications, setup instructions, and operational guide for the Quantek CP-611LSR-V2.0 Laser Photocell Sensor, featuring adjustable ranges and IP65 rating.
Praview M8/ML8 RS3 KIT Arahan Operasi Kawalan Jauh
Arahan pengendalian untuk sistem kawalan jauh M8/ML8 RS3 KIT oleh Quantek, memperincikan fungsi butang, konfigurasi dan nota keselamatan untuk operasi pintu.
Praview Quantek SQUARE & SINGLE GANG Proximity Switch Manual
Comprehensive manual for Quantek SQUARE & SINGLE GANG proximity switches, detailing installation, hardwired specifications, wiring diagrams, and radio receiver (RX-2) programming, including transmitter compatibility.