Apera Instruments AI531

Apera Instruments PH820 Benchtop pH Meter Kit

Manual Arahan

Produk Selesaiview

The Apera Instruments PH820 Benchtop pH Meter Kit is designed for reliable and accurate pH measurements in laboratory settings. It features advanced functionalities including intelligent calibration, data management, and a durable, water-resistant structure. This manual provides detailed instructions for proper use and care of your PH820 meter.

Apera Instruments PH820 Benchtop pH Meter Kit with electrode and stand

Figure 1: Apera Instruments PH820 Benchtop pH Meter Kit, showing the main unit, electrode, temperature probe, and electrode arm.

Key features of the PH820 include:

  • Ketepatan Tinggi: Achieves ±0.002 pH±1 digit precision.
  • Premium Electrode: Equipped with the Swiss LabSen 211 Glass-body pH Electrode for stable and accurate measurements.
  • Intelligent Calibration: Supports 1 to 5 points auto-calibration with calibration guide and reminder functions.
  • Diagnosis Kendiri: Assists in identifying and resolving common calibration issues and monitors electrode health.
  • Pengurusan Data: GLP-compliant data storage for up to 500 groups with USB data output and PC software support.
  • Reka bentuk tahan lasak: IP54 water-resistant construction.

Persediaan

Membongkar dan Semakan Komponen

Carefully unpack all components from the kit. Verify that the following items are present:

  • PH820 pH Meter Unit
  • LabSen 211 pH Elektrod
  • Siasatan Suhu
  • Flexible Electrode Arm and Holder
  • Kabel USB
  • PC-Link Software Flash Drive
  • Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
  • Calibration Solutions (4.00 pH, 7.00 pH, 10.01 pH)

Komponen Penyambung

Connect the electrode, temperature probe, and power adapter to the corresponding ports on the back of the PH820 meter unit. Ensure all connections are secure.

Back panel of Apera Instruments PH820 pH meter showing connection ports

Rajah 2: Belakang view of the PH820 meter, highlighting the REF, pH/mV, TEMP, RS232, and DC 9V input ports.

  1. Sambungan Elektrod: Plug the LabSen 211 pH Electrode into the port labeled "pH/mV".
  2. Sambungan Probe Suhu: Insert the temperature probe into the port labeled "TEMP".
  3. Sambungan Kuasa: Connect the power adapter to the "DC 9V" port and then to a power outlet.
  4. Electrode Arm Assembly: Attach the flexible electrode arm and holder to the side of the meter unit for convenient electrode placement during measurements.

Arahan Operasi

Basic pH Measurement

  1. Hidupkan meter dengan menekan butang kuasa.
  2. Bilas elektrod dengan air suling dan lap perlahan-lahan sehingga kering.
  3. Immerse the electrode and temperature probe into the sample solution. Ensure the electrode tip is fully submerged.
  4. Wait for the reading on the large LCD to stabilize. The meter displays pH and temperature simultaneously.
  5. Record the stable reading.

Penentukuran pH

The PH820 supports 1 to 5 points automatic calibration. Regular calibration ensures measurement accuracy.

  1. Prepare the standard calibration solutions (4.00 pH, 7.00 pH, 10.01 pH are included).
  2. Tekan ke CAL button to enter calibration mode.
  3. Rinse the electrode and immerse it into the first calibration solution (e.g., 7.00 pH).
  4. The meter will automatically recognize the solution and begin calibration. Wait for the stability icon to appear.
  5. Once stable, the meter will prompt for the next calibration point. Repeat the rinsing and immersion process for subsequent solutions.
  6. The self-diagnosis function will provide feedback on the calibration process and electrode condition, including slope data.
  7. Exit calibration mode once all desired points are completed.

Pengurusan Data

The PH820 allows for GLP data management, storing up to 500 groups of data. Data can be output via USB to a PC using the provided PC-Link software (Windows-based system only). Auto-timing data logging can also be configured.

ORP Measurement (Optional)

The PH820 can measure Oxidation-Reduction Potential (ORP) when an ORP electrode (e.g., 301Pt-C, sold separately) is connected to the pH/mV port. Refer to the ORP electrode's specific instructions for calibration and measurement procedures.

Penyelenggaraan

Penjagaan Elektrod

Proper care of the LabSen 211 pH Electrode is crucial for maintaining accuracy and extending its lifespan.

Close-up of the LabSen 211 pH electrode

Rajah 3: Jarak dekat view of the LabSen 211 pH Electrode, showing its glass body and sensor tip.

  • Bilas: Always rinse the electrode with distilled or deionized water before and after each measurement.
  • Storan: Store the electrode in its protective cap filled with a suitable electrode storage solution (or 3M KCl solution). Never store the electrode dry or in distilled water.
  • Pembersihan: If the electrode becomes contaminated or readings are sluggish, clean it according to the electrode's specific cleaning instructions.
  • Bubble-free Tip: The LabSen 211 features an impact-resistant and bubble-free sensor tip. Ensure no air bubbles are trapped around the sensor during measurement.

Penjagaan Meter

The PH820 meter unit is IP54 water-resistant, meaning it is protected from dust and splashing water. However, it should not be submerged. Keep the meter clean and dry. Use a soft, damp cloth to wipe the surface. Avoid using abrasive cleaners or solvents.

Menyelesaikan masalah

The PH820 features a self-diagnosis function that can assist in identifying and fixing common problems, particularly during calibration. Refer to the meter's display for diagnostic messages.

MasalahKemungkinan PuncaPenyelesaian
Unstable ReadingsElektrod kotor; gelembung udara; s tidak mencukupiample volume; expired calibration.Clean electrode; ensure no bubbles; use adequate sample; recalibrate.
Kegagalan PenentukuranIncorrect buffer solution; contaminated electrode; electrode aging.Use fresh, correct buffers; clean electrode; check electrode slope data (self-diagnosis) for replacement indication.
Tiada Paparan/KuasaPower adapter issue; loose connection.Check power adapter and outlet; ensure all cables are securely connected.
Bacaan Tidak TepatElectrode not calibrated; temperature not compensated; damaged electrode.Perform calibration; ensure temperature probe is functioning; inspect electrode for damage.

For persistent issues not resolved by the above, please contact Apera Instruments customer support.

Spesifikasi

ParameterNilai
Nombor ModelAI531
Ketepatan pH±0.002 pH±1 digit
Titik Kalibrasi1 to 5 points (Auto.)
Penyimpanan Data500 groups (GLP compliant)
Output DataUSB (PC software support)
Tahan AirIP54
Input Kuasa120V AC 60Hz
Output Kuasa9V 300mA
Dimensi Pakej10.67 x 10.2 x 6.65 inci
Berat badan6.69 paun
PensijilanCE, RoHS, ISO 9001:2015

Waranti dan Sokongan

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of production. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments website. You may also contact their customer service directly for assistance with product operation, troubleshooting, or parts replacement.

© 2024 Apera Instruments. All rights reserved.