This user manual provides comprehensive instructions for the safe and effective operation of the BENETECH GM-900 Infrared Thermometer. This device is designed for non-contact temperature measurement, offering a wide temperature range and various advanced features for accurate readings.
The GM-900 is suitable for measuring surface temperatures of objects that are hazardous to approach or inaccessible, such as high-voltage parts or moving machinery. Its precise laser targeting ensures accurate measurement.
Produk Selesaiview

Rajah 1: Utama view of the BENETECH GM-900 Infrared Thermometer, showing the full device and a detailed inset of its LCD display.

Rajah 2: jarak dekat view of the GM-900's LCD display, showing temperature readings, emissivity setting, battery indicator, and hold function.

Rajah 3: sebelah view of the GM-900, highlighting its ergonomic design and trigger.
The BENETECH GM-900 features a clear LCD display with backlight, intuitive button controls for settings like emissivity and unit selection, and a comfortable ergonomic grip. It includes a laser pointer for precise targeting of measurement areas.
Persediaan
1. Pemasangan Bateri
- Cari petak bateri pada pemegang termometer.
- Gently pull the handle cover downwards to open the compartment.
- Insert one 9V battery (not included) into the compartment, ensuring correct polarity (+/-).
- Tutup petak bateri dengan selamat.

Rajah 4: Battery compartment opened, showing terminals for 9V battery.
2. Kuasa Awal Hidup
After battery installation, press the trigger to power on the device. The LCD display will illuminate.
Arahan Operasi
1. Mengambil Pengukuran
- Point the thermometer at the target object. Ensure the target fills the measurement spot as much as possible. Refer to the Distance to Spot Ratio (D:S) in the Specifications section.
- Press and hold the trigger. The temperature reading will appear on the LCD display instantly.
- Release the trigger to hold the reading on the display (Data Hold Function). The "HOLD" icon will appear.
2. Unit Selection (°C/°F)
Tekan ke ° C / ° F button (usually located near the display) to switch between Celsius and Fahrenheit temperature units.
3. Pelarasan Emisiviti
Emissivity (EMS) is crucial for accurate infrared temperature measurement. The GM-900 allows adjustment from 0.10 to 1.00 (default is 0.95).
- Tekan ke SET button to enter emissivity adjustment mode.
- Use the Up/Down arrow buttons to increase or decrease the emissivity value.
- Tekan SET sekali lagi untuk mengesahkan dan keluar.
4. Penunjuk Laser
The laser pointer assists in targeting the measurement area. Press the laser button (often depicted with a laser icon) to toggle the laser on or off.
5. Lampu latar
Press the backlight button (often depicted with a light bulb icon) to toggle the LCD backlight on or off for better visibility in different lighting conditions.
6. Advanced Functions (MODE Button)
The MODE button cycles through various measurement functions:
- MAKS: Memaparkan suhu maksimum yang direkodkan semasa imbasan berterusan.
- MIN: Memaparkan suhu minimum yang direkodkan semasa imbasan berterusan.
- AVG: Memaparkan suhu purata yang direkodkan semasa imbasan berterusan.
- DIF: Displays the difference between the maximum and minimum temperatures recorded.
- HAL (Penggera Tinggi): Set a high temperature alarm. If the measured temperature exceeds this value, an alarm will sound.
- LAL (Penggera Rendah): Set a low temperature alarm. If the measured temperature falls below this value, an alarm will sound.
- STO (Storage): Access memory storage (up to 12 data points).
To set alarm values, select HAL or LAL mode, then use the Up/Down arrow buttons to adjust the threshold. Press SET untuk mengesahkan.
Penyelenggaraan
1. Membersihkan Lensa
The infrared lens is the most delicate part of the thermometer. Clean it carefully with a soft cloth or cotton swab and water, or medical alcohol. Do not use abrasive cleaners or solvents.
2. Membersihkan Casing
Clean the thermometer casing dengan iklanamp cloth and mild soap. Do not immerse the device in water.
3. Penyimpanan
Store the thermometer in a cool, dry place, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the battery to prevent leakage.
Menyelesaikan masalah
| Masalah | Kemungkinan Punca | Penyelesaian |
|---|---|---|
| Peranti tidak dihidupkan. | Bateri rendah atau mati; pemasangan bateri yang salah. | Replace the 9V battery. Ensure correct polarity. |
| Bacaan yang tidak tepat. | Incorrect emissivity setting; lens is dirty; target too far or too small. | Adjust emissivity for the target material. Clean the lens. Ensure target fills the measurement spot. |
| Laser pointer not working. | Fungsi laser dimatikan. | Press the laser button to activate. |
| Paparan malap atau berkelip. | Bateri lemah. | Ganti bateri 9V. |
Spesifikasi
| Parameter | Nilai |
|---|---|
| Julat Suhu | -50 ~ 950 ° C (-58 ~ 1742 ° F) |
| Ketepatan | ±1.5 °C or ±1.5% (0°C to 950°C); ±3 °C (-50°C to 0°C) |
| Resolusi | 0.1 ° C atau 0.1 ° F |
| Kebolehulangan | 1 % or 1 °C |
| Masa Tindak Balas | 500 ms (0.5 seconds), 95% response |
| Tindak Balas Spektrum (Panjang Gelombang) | 8 ~ 14 µm |
| Nisbah Jarak ke Titik (D:S) | 12:1 |
| Emisiviti | 0.10 ~ 1.00 (Adjustable, Preset at 0.95) |
| Suhu Operasi | 0 ~ 40 ° C (32 ~ 104 ° F) |
| Suhu Penyimpanan | -20 ~ 60 °C (-4 hingga 140 °F) |
| Kelembapan Relatif | 10 ~ 95% R.H. non-condensing, up to 30 °C |
| kuasa | 1 x Bateri 9V (tidak termasuk) |
| Hayat Bateri | 22 hours (Non-Laser Mode), 12 hours (Laser Mode) |
| Dimensi (L x W x D) | approx. 175 x 100 x 49 mm (6.89 x 3.94 x 1.93 inch) |
| Berat (tanpa bateri) | lebih kurang 226 g (7.97 oz) |
Waranti dan Sokongan
For warranty information and technical support, please refer to the documentation provided with your purchase or contact BENETECH customer service. Keep your purchase receipt as proof of purchase.
For further assistance, please visit the official BENETECH webtapak atau hubungi pusat perkhidmatan mereka yang dibenarkan.